Instrument setup:
The following instruments were provided by Brian Hogue's mooring shop:
SBE37SM-RS232 serial no. 6662, deployed on the top sphere at 33m
SBE37SM-RS232 serial no. 6661, deployed on the wire at 50m;
SBE37SM-RS232 serial no. 14642, deployed on the wire at 67m
SBE37SM-RS232 serial no. 14639, deployed on the wire at 81m
SBE37SM-RS232 serial no. 14640, deployed on the wire at 95m
SBE37SM-RS232 serial no. 14633, deployed on the wire at 109m
SBE37SM-RS232 serial no. 14637, deployed on the wire at 123m
SBE37SM-RS232 serial no. 6667, deployed on the ADCP cage at 135m
They were configured to record temperature, conductivity, and pressure
at 15-minute intervals.
The data were processed by Adam Houk. The following text as well as
the above two figures were taken from his
instrumentation notes:
"A total of 8 SeaBird SBE37-SM microcats were installed on the BS-3
mooring: one on the top buoy, and 7 at various points along the mooring.
All 8 instruments were sampling upon recovery and had full and complete
records, with minimal clock drift. With one exception, all the microcat
data appear to be reasonable. Serial number 14637, which was positioned
at 123 meters, recorded a nominal pressure value of about 93 decibars.
The position of the instrument was verified, and there was no sign of
significant pressure drift in the data, which suggests that its pressure
sensor is faulty. The instrument should be sent to SeaBird for inspection/repair
before being used again. "
|