Deployment summary: Duration: Oct 25, 2020 to Nov 4, 2022 Location: 71° 23.667' N, 152° 3.080' W in 147 meters of water Variables measured: chlorophyll fluorescence, nephelometric turbidity Sensor information: FLNTU serial no. 4398; for details see the manual; Instrument settings: sampling interval 1 minute Calibration: based on factory calibration data: contained in the matlab file bs3_flntu_2020-22.mat. We also include here the downloaded raw data file bs3-02_fluorometer_raw_clipped.raw . Data processing: The data were processed by Adam Houk. He found the instrument clock to be 21 minutes slow and applied a linear drift over the course of the deployment to correct for it. The conversion from raw data to chlorophyll concentration adnd turbidity followed the formula on the calibration sheet. Adam noted that third raw data channel did not produce a realistic temperature when applying a conversion formula that had "worked" for the previous deployment: (temp = (temp_raw-574) / -2.5). However, according to the manual, "the "Therm" column is NOT a calibrated signal. It is used for factory diagnostics, but does not have a physical temperature characterization performed. It is an on-board (internal) CPU chip thermistor signal. It varies inversely to the internal temperature of the ECO, between 480 - 600 counts (~530 for room temperature, 23C). " Since the raw data were below 600 here, we did not pursue this further. |
|
|