Deployment 2020 to 2022:
FLNTU

Figure provided by Adam Houk.
Deployment summary:

Duration: Oct 25, 2020 to Nov 4, 2022
Location: 71° 23.667' N, 152° 3.080' W in 147 meters of water
Variables measured: chlorophyll fluorescence, nephelometric turbidity
Sensor information: FLNTU serial no. 4398; for details see the manual;
Instrument settings: sampling interval 1 minute
Calibration: based on factory calibration
data: contained in the matlab file bs3_flntu_2020-22.mat.
We also include here the downloaded raw data file bs3-02_fluorometer_raw_clipped.raw .


Data processing:

The data were processed by Adam Houk. He found the instrument clock to
be 21 minutes slow and applied a linear drift over the course of the
deployment to correct for it. The conversion from raw data to chlorophyll
concentration adnd turbidity followed the formula on the calibration
sheet.

Adam noted that third raw data channel did not produce a realistic
temperature when applying a conversion formula that had "worked" for the
previous deployment: (temp = (temp_raw-574) / -2.5). However, according
to the manual, "the "Therm" column is NOT a calibrated signal. It is
used for factory diagnostics, but does not have a physical temperature
characterization performed. It is an on-board (internal) CPU chip
thermistor signal. It varies inversely to the internal temperature of the
ECO, between 480 - 600 counts (~530 for room temperature, 23C).
"
Since the raw data were below 600 here, we did not pursue this further.