Deployment 2016 to 2018
temperature and conductivity calibration


The calibration adjustments were based on the drift determined from the laboratory
calibrations at SeaBird before and after the deployment. The SeaBird sheets, while
very helpful, can be confusing at first sight (they have confused me in the past).
Below, the adjustment process is described for the example of microcat #2139.
It is followed by a summary table with the parameters for all eight microcats,
and the SeaBird calibration figures.


Temperature:
The plot below shows the lower portion of the SeaBird calibration sheet for the March/April 2019
post-cruise calibration, a graphical display of the residuals from the current and the most
recent previous calibration. When comparing calibrations from different times, the SeaBird
sheets use the current (2019) coefficients to convert current instrument and bath readings
(the triangles close to the zero line) as well as the earlier instrument and bath readings
(circles). This general guide was described by SeaBird's Mark Baumann in an email from
June 2000:

      "In general all comparisons on the cal sheet, be it temperature or
      conductivity are done using the coefficients for the calibration date that
      the coefficients are calculated for (the date at the upper left corner of
      the cal sheet) and the sensor outputs from the other calibrations (the
      dates to the right of the residual plot)."

Residuals are defined as instrument temperature minus bath temperature.



The figure indicates a small drift of about 0.002 degrees C over the range of values
encountered during deployment (0 to 7 deg C). The cell has drifted [very slightly] "low"
: using the 2016 instrument readings with the 2019 cal, instrument temperature was larger
(residuals positive) than bath temperature. This implies that converting sensor frequencies
collected in 2019 into temperatures using the 2016 cal generates instrument readings that
are too low.


A simple example:
Lets say: temperature = cal * sensor frequency, a simplified conversion relation;
- the 2019 sensor frequency = 5; bath temp = 10; => 2019 cal = 2;
- we know: 2016 frequencies and 2019 cal generated positive residuals,
      i.e., higher temps; lets say 15 degrees; => 2016 frequency was 7.5;
- back in 2016, a frequency of 7.5 gave the correct bath temp of 10, so the
      2016 cal was 1.33.
- in 2019, when the sensor frequency was 5, applying the 2016 cal gives only 6.67 degrees
      => we have to increase the temps calculated by converting frequencies measured
      with the 2016 cal.

Since we are using the 2016 cal (i.e., the T values downloaded from the instrument
after recovery), we need to add 0.002 deg C at the end of the deployment.
Lacking additional information, we assume the drift occurred linearly in time over
the course of the deployment, from no correction at the start to full correction at
the end of the record.




Conductivity:
The conductivity correction followed a similar scheme. Again, the sensor had drifted "low",
leading to positive residuals on the cal sheet. So conductivity of the downloaded data was
increased in time, again going linearly from no to full correction over the course of the
record. In contrast to temperature, however, the correction itself was a linear funtion of
conductivity ("slope correction"). The reason is that the origin (zero output at zero
conductivity) has been found to be very stable.





Pressure:
The pressure sensor tends to be quite stable. This was confirmed by the
SeaBird post-deployment calibration (see below), and no adjustment was
needed for any of the microcats.



Summary info for all microcats

All calibration adjustments were small. The list below summarizes conductivity slope
and temperature bias corrections. Both varied linearly from no to full correction over
the course of the deployment. At the time of recovery,

      T_corrected = T_recovered + T_offset;
      C_corrected = C_recovered * C_slope;

with the following values :

serial # depth T offset C slope
14534 33m 0.0007 deg 1/1 for now!
14535 52m 0.0 deg 1/0.9994780
14551 133.5m 0.0017 deg 1/0.9992461
14552 66.5m 0.0022 deg 1/0.9988205
14553 79.5m 0.0028 deg 1/0.9990221
14554 94m 0.002 deg 1/0.9991847
14555 107.5m 0.001 deg 1/0.9994190
14556 121m 0.0 deg 1/0.9994781



Below, the SeaBird T and C calibration sheets as well as the adjustment plots
are shown in one tow of plots per microcat.