The 2013-14 MC TS (added as slightly darker dots below) follows more closely the earlier CMPs rather than the 2013-14 record. |
|
|
|
Density profiles also suggest low 2013-14 values. Look at periods of little vertical gradient: the appear to be at lighter densities compared to earlier years. |
|
|
|
The following plots step through our method of comparing deep CMP bins to the MC record. It involves estimating the vertical gradient from the CMP record to extrapolate the CMP to the MC depth level. As a first step, I look at conductivity profiles (and also temperature, as explained in a moment) to identify general structure. The conductivity figure suggests a strong gradient region between 60 and 80 dbar during the earlier profiles (blue-ish colors). In the following, I chose depths below 80 dbar for calculating a vertical gradient. Although we do not calibrate T, I start by illustrating how the calibration method works by looking at T. For consistancy, I also calculated vertical T gradients from depths below 80 dbar. |
|
|
|
The vertical temperature gradient was then determined by linear fit for all profiles with at least 4 points below 80 dbar (below left, top panel). The quality of the fit was roughly estimated by the mean of the absolute difference between fit and profile (bottom), with gradients below a "quality threshold value" marked in green in the top panel. Gradients were variable early in the deployment, while being close to zero in later portion. The CMP temperature at the depth of the microcat was then derived from extrapolating the linear fit to the deep CMP bins down to the microcat pressure. The top panel of the right figure below shows both MC (blue) and extrapolated CMP (red) temperatures, while the bottom panel shows their difference,with linear fits below the quality threshold value marked in black (sorry, green would have been more consistent). |
|
|
|
Another way to look at this issue, presented below left, directly compared the CMP and microcat measurements (i.e., leave out the gradient calculation and extrapolation): in the left figur ebelow, the MC temperature time series is shown in black, while the deepest eleven CMP bins were added in color ranging from blue (116 dbar) to red (136 dbar). Particularly the profiles with little vertical CMP spread showed good agreement with the MC profile. In the most complicated of my figures, below right, the MC time series was plotted only for those times for which deep CMP bins were available (again in black). In essence, its a zoom-in to the times where we can compare CMP and MC. I also added the extrapolated CMP temperature at the MC depth level (red line). Both lines match most of the CMP clusters for the second half of the record where the vertical gradient was less noisy. |
|
|
|
Repeating the same approach for conductivity produced results that at first appeared similar: The conductivity gradients showed less spread during the later record, though in this case averaged around non-zero value (C increasing with depth, below left). |
|
|
|
When comparing the full MC and CMP conductivity records (below left), the CMP clusters (colored stars) appear not too far off the MC curve (black line). Similarly, when looking only at the times of deep CMP data (below right), things appear to make sense: e.g., for the last set around decimal day 545, CMP colored stars increase as their colors change from blue (shallower, farther away from the MC) to red (closest to the MC). After applying the estimate of the vertical C gradient, the extrapolated CMP values (red curve) fall short of the MC line, i.e., CMP conductivity appears to be too low. That is consistent with the impression from the TS diagrams shown at the very top of this note. The problem is in the magnitude of that shortfall. When expressed in units of mS/m, conductivity changes correspond roughly 1:1 to salinity changes; e.g., the 0.2 mS/cm grid of the figure in the lower left corresponds to about 0.25 psu at these temperatures and pressures. So jumping back for a moment to the figure ABOVE right, where the bottom panel shows the difference between extrapolated CMP conductivity and MC conductivity, even the more "quiet" second half of the record implies MC-CMP salinity differences between 0 and 1.2 psu. In summary: We have strong evidence of a significant conductivity difference at the end of the record, few datapoints to quantify its magnitude, and even less information at the beginning as to whether this was a drift or an offset. |
|
|