conductivity calibration concerns

The 2013-14 CMP had profiling issues. Due to a firmware bug, odd-numbered profiles
were not saved; thus only climb profiles (bottom to top crawl stops) were available. In
addition, the profiler was ballasted slightly light, which caused it to drift towards
the top stop in the six hours between profiles. As a result, many profiles were shortened
or simply recorded only at the top stop.

A plot of all available depth bins below shows the extend of the loss: overall, about 18%
of all possible climb data were occupied (i.e., only 9% when considering the missed dive
profiles as well), and was worse for the 10 deepest bins with only 2% coverage there.



We calibrate the CMP conductivity by comparing its deepest bins to the microcat
immediately below the bottom stop. For the above reasons, few data points were available
for the cal: while a complete record would have about 1000 points to compare, we have
here around 20.

For a first look, we compare the TS diagram (below) with those of earlier deployment.
It appears to me that the 2013-14 salinities were low, perhaps by slightly more than
0.5 psu compared to 2011-12 (left) and 2012-13 (right). Note that color indicates time
from blue (early, starting at deployment) to red (late, ending at recovery).

The 2013-14 MC TS (added as slightly darker dots below) follows more
closely the earlier CMPs rather than the 2013-14 record.
Density profiles also suggest low 2013-14 values. Look at periods of little
vertical gradient: the appear to be at lighter densities compared to earlier
years.
The following plots step through our method of comparing deep CMP bins
to the MC record. It involves estimating the vertical gradient from
the CMP record to extrapolate the CMP to the MC depth level.

As a first step, I look at conductivity profiles (and also temperature,
as explained in a moment) to identify general structure. The conductivity
figure suggests a strong gradient region between 60 and 80 dbar during
the earlier profiles (blue-ish colors). In the following, I chose depths
below 80 dbar for calculating a vertical gradient.

Although we do not calibrate T, I start by illustrating how the calibration
method works by looking at T. For consistancy, I also calculated vertical
T gradients from depths below 80 dbar.

The vertical temperature gradient was then determined by linear fit for
all profiles with at least 4 points below 80 dbar (below left, top panel).
The quality of the fit was roughly estimated by the mean of the absolute
difference between fit and profile (bottom), with gradients below a
"quality threshold value" marked in green in the top panel.

Gradients were variable early in the deployment, while being close to zero
in later portion.

The CMP temperature at the depth of the microcat was then derived from
extrapolating the linear fit to the deep CMP bins down to the microcat
pressure. The top panel of the right figure below shows both MC (blue)
and extrapolated CMP (red) temperatures, while the bottom panel shows
their difference,with linear fits below the quality threshold value
marked in black (sorry, green would have been more consistent).

Another way to look at this issue, presented below left, directly compared
the CMP and microcat measurements (i.e., leave out the gradient calculation
and extrapolation): in the left figur ebelow, the MC temperature time
series is shown in black, while the deepest eleven CMP bins were added
in color ranging from blue (116 dbar) to red (136 dbar). Particularly
the profiles with little vertical CMP spread showed good agreement with
the MC profile.

In the most complicated of my figures, below right, the MC time series
was plotted only for those times for which deep CMP bins were available
(again in black). In essence, its a zoom-in to the times where we can
compare CMP and MC. I also added the extrapolated CMP temperature at the
MC depth level (red line). Both lines match most of the CMP clusters for
the second half of the record where the vertical gradient was less noisy.

Repeating the same approach for conductivity produced results that at first
appeared similar: The conductivity gradients showed less spread during the
later record, though in this case averaged around non-zero value (C increasing
with depth, below left).

When comparing the full MC and CMP conductivity records (below left),
the CMP clusters (colored stars) appear not too far off the MC curve
(black line). Similarly, when looking only at the times of deep CMP data
(below right), things appear to make sense: e.g., for the last set
around decimal day 545, CMP colored stars increase as their colors
change from blue (shallower, farther away from the MC) to red (closest
to the MC). After applying the estimate of the vertical C gradient,
the extrapolated CMP values (red curve) fall short of the MC line,
i.e., CMP conductivity appears to be too low. That is consistent
with the impression from the TS diagrams shown at the very top of
this note.

The problem is in the magnitude of that shortfall.

When expressed in units of mS/m, conductivity changes correspond
roughly 1:1 to salinity changes; e.g., the 0.2 mS/cm grid of the
figure in the lower left corresponds to about 0.25 psu at these
temperatures and pressures.

So jumping back for a moment to the figure ABOVE right, where
the bottom panel shows the difference between extrapolated CMP
conductivity and MC conductivity, even the more "quiet" second
half of the record implies MC-CMP salinity differences
between 0 and 1.2 psu.

In summary: We have strong evidence of a significant conductivity
difference at the end of the record, few datapoints to quantify
its magnitude, and even less information at the beginning as to
whether this was a drift or an offset.