Hi Frank, Here are the recommended settings for the BPR 53-7: tide measurement: interval = 15 minutes, duration = 2 minutes, power pressure sensor continuously measure reference frequency every 1440 tide samples --------------------------------------------------------------------------------------- Here is the recommended deployment configuration for IPS4 1043 (cfg file attached): [Deployment] DeployName = "bs3x12" NumPhases = 1 SerialNumber = "1043" [phase01] year = 12 month = 8 day = 13 hour = 0 ping_interval = 2 ping_length = 68 lockout = 2 max_range = 60 pressure_interval = 20 burst_interval = 510 burst_count = 1 record_persistence = 0 record_amplitude = 1 min_persistence = 8 start_amplitude = 200 stop_amplitude = 150 Rick --- Richard Krishfield Woods Hole Oceanographic Institution Clark 128 - MS 23 Woods Hole, MA 02543-1541 USA Phone: (508) 289-2849 Fax: (508) 457-2163 e-mail: rkrishfield@whoi.edu