BS3 2012-13
conductivity calibration

We calibrate the CMP conductivity by comparing its deepest bins to a microcat (MC)
that is typically located immediately below the bottom stop. In this deployment,
however, there was a PCO2 sensor between the two, increasing the vertical
separation between the CMP and MC. Also, additional floatation had been added
below the MC, so that the CMP bottom stop as well as the MC were located slightly
higher in the water column than in previous years.

The figure below illustrates the overall vertical coverage of the CMP, as well
as the depth location of the deepest bins.



The CMP conductivity profile time series included periods of strong gradients
early and late in the deployment, as well as periods of small and nearly
constant gradients; colors below indicate time from early (blue) to late (red).



The main difficulty for the calibration of this dataset was the determination
of the conductivity gradient to be used for extrapolating CMP conductivity
down to the MC depth. Based on the above figure and some tests, we settled
on the depth range fron 116 to 124 dbar, or the five deepest CMP bins.



With the gradient determined, we can extrapolate the CMP conductivity down
to MC depth, and (after interpolating the MC onto the CMP time grid) compare
the two. The top panel below includes conductivity from the MC (red), the CMP
extrapolated to MC depth (black), as well as the deepest five bins of the
CMP (colored stars). Large fluctuations seen early and late in the
deployment coincided with periods of strong vertical gradient (e.g.,
around dday 350, 2012/12/16), with occasional "quieter" periods interspersed
where the plotted values closely matched also occurred, with the longest one
lasting about 50 days (dday 500 to 550, mid-May to the start of July).



With the conductivity origin (0 S/m for fresh water) assumed to be
unchanging, conductivity calibrations are traditionally applied as scale
factors, derived here as the ratio of correct (MC) over measured (CMP) values.
Using an iterative approach copied from Paula Fratantoni's CMP software,
we first excluded the largest ratio outliers, then successively removed
outliers around a linear fit of the remaining values. The final fit
(magenta line in the bottom panel above) ranged from 1.00015 to 1.00123
over the course of the year-long deployment, corresponding to a salinity
adjustment of 0.005 to 0.045 psu.

The effect of the adjustment is shown in a couple ways below. First, we
compare the time series in the previous figure, zooming in on three times
where vertical gradients were small from before (first row of plots below)
and after (second row). While these figures appear noisy, reflecting the
variability of the dataset and the increased vertical separation between
CMP and MC relative to earlier years, the red and black curves (MC and
extrapolated CMP, respectively) are closer following the adjustment.






We also compare CMP and MC in TS space before and after applying the
calibration. Although the MC data (light grey underneath the CMP data
colored to indicate profile time from early in blue to late in red)
have a larger spread than the fewer deep CMP data, the conductivity
adjustment appears to better center the CMP TS curves on the MC spread.