Deployment 2010 to 2011:
Shallow Microcat "SBE37"




Deployment summary:
Duration: Sept 16, 2010 to Oct 11, 2011
Location: 71° 23.650' N, 152° 2.810' W
Variables measured: temperature, conductivity, pressure
Sensor information: SBE37-SM V 2.6b SERIAL NO. 2134 ;
Instrument settings: sampling interval 15 minutes
Calibration: based on pre- and post cruise SeaBird laboratory calibration detailed here
Processed data: contained in the matlab file bs3_201011_2134_cal.mat


Instrument setup and recovery:
The instrument was provided by Scott Worrilow's mooring shop, with calibrations routinely performed at SeaBird's lab facilities. Onboard the Healy shortly before deployment, the instrument was opened, the lithium battery pack was connected, and the instrument was connected via an RS485<->RS232 converter to a PC. For communications, I used Seabird's Seaterm program, checking that "configure" was set for "sbe37 rs485". The instrument clock was checked against the ship's UTC time server, and the standard data collection parameters were confirmed; communications with the instrument were captured in a log file.

The mooring was recovered on Oct 11, 2011. On the 12th, I connected to the SBE37 (Seaterm log file attached). A clock check indicated the instrument clock to be 1 minutes 49 seconds fast:


Data processing:
Data processing consisted of the clock adjustment as well as adjustments to the temperature and conductivity records based on a post-deployment calibration at Seabird. It had determined a 1 mdeg C temperature drift for out measured temperature range, and a conductivity slope adjustment that amounted to about 0.0016 S/m, or about 0.016 psu for the observed conductivity range (sensor read high). The SBE37 record was adjusted assuming a linear drift from 0 at the start of the record to the post-deployment values at sensor recovery. More details of the calibration adjustment can be found here.

Lastly, the data were inspected for sensor glitches, but none were taken out (what appears like pressure spikes in the 1-year plot were in fact events with some duration).

.