The calibration adjustments were based on the drift determined from the laboratory
calibrations at SeaBird before and after the deployment. The SeaBird sheets, while
very helpful, can be confusing at first sight (they have confused me in the past).
Temperature:
The plot below shows the lower portion of the SeaBird calibration sheet for the Nov 5, 2009
post-cruise calibration. The triangles close to the zero line represent the residuals from
fitting bath temperatures to SBE37 output using the new (2009) calibration values. The
circles indicate the residuals when using the same 2009 measurements, but converting the SBE37
raw frequency output to temperatures via the calibration coefficients determined
back on March 26, 2008. In other words: without recalibrating the sensor (i.e., leaving it
with the 2008 cal), in January of 2011 the SBE37 measured temperature too high. SeaBird
determined the offset to be 0.00103 ° C when averaged over the whole calibration range.
However, for the temperature range we obeserved in the field around zero ° C, it was
about 0.0027 ° C.
Without further knowledge, the simplest model would be to assume that the sensor drifted linearly in time.
It is further generally assumed that the drift occurs primarily during sensor deployment, when
the sensor was actually in the ocean. There are then two possible approaches for applying a correction:
1) use the 2009 calibration coefficients to re-generate temperatures from sensor frequencies contained in the raw
data file, presumably via SeaBird programs; then adjust linearly from maximum
correction at the deployment start (when the 2008 calibration was correct) to zero correction
at the end (when the 2009 calibration was correct).
2) start with the existing temperature data file generated with the 2008 calibration, and adjust it
linearly from zero correction at deployment to maximum correction at recovery.
For simplicity, we chose the second approach.
To correct our field measurements, we therefore subtracted an offset that varied linearly
from zero upon deployment to 0.00027 ° C on recovery. The plot below displays the correction
applied to the temperature record.
Note: the SeaBird calibration plot above suggests that the calibration change was not constant over the range of bath
temperatures, i.e., a "slope correction" rather than a constant offset may be more appropriate. Still, the
correction suggested by SeaBird was a constant offset, which is standard practice for their temperature
probes. This seems reasonable here given the small temperature range we encountered in the field.
Conductivity:
The conductivity correction followed a similar scheme. Again, the first plot below shows the
figure portion of the SeaBird post-cruise calibration sheet. In this case, using the March 2008
calibration coefficients showed residuals for the 2009 calibration measurements to be positive:
i.e., the conductivity in the data file generated with the pre-cruise calibration was too high
at the end of the deployment. In contrast to temperature, however, the correction itself was
a linear funtion of conductivity ("slope correction").
As for temperature, we chose to use the 2008 cal to convert sensor frequencies into engineering data (conductivity),
then linearly adjusted from zero slope correction at deployment time to full post-cruise calibration slope
correction on recovery. The actual correction in terms of S/m subtracted from the original record
(shown again in the lower panel of the plot below) varied roughly linearly from 0 to 0.00038 S/m over
the time the sensor was in the ocean. It was of course close to zero at the very end when the
sensor was back on deck but not yet turned off, as conductivity itself was zero then.
Pressure:
This unit did not have a pressure sensor.
|