Deployment 2008 to 2009:
Arctic Winch (AW)

final data


Deployment summary:
Duration: Aug 12, 2008 to July 29, 2019
Location: BS3 mooring site
Variables measured: pressure, temperature, conductivity, fluorescence, turbidity
Sensor information: The Arctic Winch is a WHOI-developed instrument platform. For this
    deployment, it was equipped with an RBR CTD.
Instrument settings: output 6 Hz data
Calibration: see text below
Processed data: contained in the matlab files aw_bs3_200809_final.mat. The file format is described here.



Data processing:

Processing steps include:
- editing;
- conversion of temperature/conductivity to salinity and density;
- lagging of T/C relation to reduce spiking and inversions;
- pressure binning into standard profiles;
- conductivity calibration.
- conversion of fluorescence/backscatter voltages;


Editing

Basic range editing provided a first filter, excluding points beyond simple limits
of -10 to 200 dbar, -5 to 15 deg C, and 15 to 50 conductivity units, respectively.
For de-spiking, number of different approaches had been explored for previous dataset,
eventually settling on acomparison of consecutive time series segments to a
second-order fit. Outliers were identified as exceeding n standard deviations.
A minimum outlier size was set to avoid editing of good points during very low
variability segments. After some experimenting, I selected a subset size of 50 points,
and a spike threshold of 2 std. I set a minimum spike size of 1 dbar, 0.1 deg C,
and 0.5 conductivity units.


Sensor lags

The Arctic winch CTD rests at depth, then ascents to the surface once a day. I split
the daily profile into up (ascent) and down (descent) profiles, with the latter
quite clearly providing more realistic data. The physical reason is presumably
the CTD location below the float.

Salinity and density were derived for several lags of temperature versus conductivity
(-3 to +2 scans). A best lag was selected based on visual inspection of the down profiles.
A -2 scan lag was selected for all but one profile here. For context, previous RBR datasets
had been processed with a lag of 0 and -2, while +1 had been used for FSI CTD datas earlier.


Vertical binning

Next, the down cast was pressure binned, using a bin size (0.5 decibar) to match previous
Arctic Winch datasets. The average profiles were screened once more for excessive density <
inversions. Three profiles were edited to remove salinity and density for a total of five bins.


Calibration

We typically use a nearby SeaBird SBE37 "microcat" record to check the Arctic Winch
conductivity calibration. Unfortunately, such a record was not available here. Instead,
we looked at the MMP data further below. As a first test, we plotted the TS relations
of both Arctic winch and MMP.

The plots below show MMP TS in red and AW TS in blue. The microcat at the deep profiling
end of the MMP was added in black for further context. The left and middle figure below
show the same data in different plot order (i.e., "who's on top"). Variability increased
towards the the surface, so while similar patterns were fairly obvious between MMP and
microcat at depth, they were less easily found between AW and MMP further up in the water
column. As a first guess, AW salinity was adjusted by 0.39 psu (right figure) as that
seemed to better align the two TS records.

To compare AW and MMP, we first estimated the vertical conductivity gradient of the
shallowest portion of the MMP record. Selecting the top four two-meter bins located
beween between 54 and 60 meters, we fitted a linear conductivity gradient and summed
the absolute difference to the observed data.

To compare AW and MMP, we first estimated the vertical conductivity gradient of the
shallowest portion of the MMP record. After fitting a constant gradient to the top four
two-meter bins (located beween between 54 and 60 meters), we selected profiles with
small residuals between fit and observed data to improve confidence that the determined
gradient could be extended updward towards the AW depths.

The top planel of the figure below shows bin the AW bin 79 centered on 39.25m in red,
and the extrapolated MMP conductivity in blue (the figure did not change significantly
for AW bin 80, not shown). The bottom panel shows the difference between the two in
cyan (in blue when using AW bin 80). The two panels suggested that there were time
periods where properties observed by the two instruments differed more strongly, e.g.,
around decimal day 330. After manually excluding these periods, we calculated a linear
fit (black line) over the remaining MMP-AW conductivity differences (black stars).



It provided a conductivity adjustment between 0.16 mS/cm and 0.3mS/cm between the start
and end of the record, corresponding to salinities of 0.20 and 0.38 psu (using T=1 deg C
for this estimate).


A/D sensor conversion

The Arctic winch sensor suite included a WetLabs FLNTU fluorescence / backscatter sensor.
Sensor voltages were converted to chlorophyll in micrograms/liter (ug/l) and backscatter
in nephelometric turbidity units (NTU) based on the manufacturer's "characterization
sheets" (calibration sheets). Unfortunately the serial number of the instrument used on this
deployment is not known. Instead, we rely on typical conversion numbers (e.g., see the
calibration sheets from FLTNU's # 2674 and 2688 that were part of the instrument pool used
for the Arctic Winches. The factory calibration coefficients differ little from instrument
to instrument. It is our understanding that a precise calibration depends more on local
species compositions than on variations amongst instrument components as reflected in
the calibration sheets. Since we do not have species compositions, our "generic" cal
coefficients seem sufficient here.

We used the following conversions:
      fluorescence [ug/l] = 10 * (fluorescence voltage - 0.08)
      turbidity [NTU] = 5 * (turbidity voltage - 0.056)